A one-page reference for the Temp Monitor Lite. For full specifications, see Temp Monitor Lite specifications.
At a glance
| Placement | Inside the Storage Unit, or in a Room |
| Probes | None; the sensor is built in |
| Monitors | One Storage Unit or one Room |
| Display | No |
| Button | One, used only to power the device on (long press, 5 seconds) |
| Local alert | Red LED only; no audible alarm |
| Installation | Kio Setup Manager |
| Calibration | Can be ordered NIST-certified |
LED indicators
The LED is the only indicator on the Temp Monitor Lite. Use it to check the device when it is installed inside a Storage Unit.
| LED | Meaning | Status in Temperature Monitoring |
|---|---|---|
| Blinking green | Monitoring normally; no active alert | Online |
| Blinking red | Active alert or hardware error | Excursion or Device error |
| Blinking yellow | Battery low; no active alert | Low battery |
| Not blinking | Paused, in the startup period after install or restart, or just powered on | Paused, Syncing, or Not reporting |
Note: If the LED is not blinking and the Monitor is not intentionally paused, check the device status in Temperature Monitoring.
Local alert behavior
The Temp Monitor Lite signals an active excursion or hardware error with a blinking red LED. It has no audible alarm. The LED is the only local indicator, and the red LED clears only when both the alert is resolved in Temperature Monitoring and the reading is back in range. For a local audible alarm or an on-device display, use the Temp Monitor Pro.
Installation notes
- Place the Monitor inside the Storage Unit, or in the Room being monitored, so the internal sensor samples the environment directly.
- The Temp Monitor Lite has no probe ports and cannot use external probes.
- Assign a Threshold Setting during installation in Kio Setup Manager; update it later in Temperature Monitoring.